Operating the GBC 902
Atomic Absorption Spectrometer


Part 1: Setting Up the Instrument


David L. Zellmer
Department of Chemistry
California State University, Fresno
Fall 1996

WARNING!
Operation of this instrument involves the use of potentially explosive gases. DO NOT operate this instrument unless directly supervised by a faculty member or trained member of the staff. The directions below contain a great deal of safety information, but are not meant as a substitute for direct supervision.

Before turning the power on to the instrument, there are several safety checks that must be made.

The hood must be on.

The trap must be full of liquid.

The flow controls for fuel and air must be off.

The acetylene tank must be either off or set to a safe operating pressure. (See below.)

Wear your safety glasses!


The switch for the hood is located at the base of the walk-in hood to the left of the instrument. Turn the hood on before doing any work with the instrument.

The flow controls for fuel and air should be at their full clockwise (off) position. These are needle valves, so turn them gently clockwise until you feel them seat. Forcing them may cause damage to the valves.

The acetylene tank should have been left off by the last person to use it. Verify this by noting that both gauges read zero. Verify that the main valve for the tank is off by attempting to turn the tank valve to the right as shown.

Verify that the pressure to the instrument is indeed set to zero by feeling the regulator control. It should rotate loosely when in the zero pressure position. Later on you will apply pressure to the system by turning this control clockwise, but for now leave it in the "loose" zero pressure postition.

Select the type of lamp you will need for your analysis. The lamp shown is a multi-element lamp, good for Fe, Co, Ni, Mn, Cu and Cr.

Also note the maximum operating current for the lamp. We generally will operate at a current much lower than the 20 mA shown on the label to conserve lamp life.


The lamp is inserted all the way into the lamp port. There are two sockets that can be attached to the lamp. Note which one you have selected.

After we turn the instrument on, we will be able to use the two lamp position controls to maximize the light sent through the instrument.

What looks like a second lamp port to the left is the permanently installed background corrector lamp.


Be sure the lamp is pushed all the way in.

It is now safe to turn the power on. The instrument should be allowed to warm up for 20 minutes before critical measurements are made. You can proceed with the set-up immediately, however.

There is a chart on the wall to the right of the instrument that gives operating parameters and information on the analysis you plan to do.

In this case:

Fe is the Element

5 is the lamp current in ma

248.3 is the wavelength in nm

0.2 is the recommended slit width in nm

0.05 - 20 is the recommended concentration range for this element in mg/liter.

(Your results can be quite different from this concentration range, so be prepared to change your standards after running them.)


Send power to the lamp by first pressing the Lamp Selection button. Be sure you have selected the correct lamp.

Use the numeric keypad to enter the current. In this case 5 ma has been entered.

Send the 5 ma to the lamp by pressing the Read button.

The "25 ma" seen in the lower part of the display is just the instrument giving you the possible range for input.


If you aren't sure the correct lamp got turned on, take it out of the lamp port and look at it. The orange glow tells you it is working properly.

Since these lamps give off UV radiation, do not point them directly into your eyes.


Set the slit control to the value recommended on the chart. Here 0.2 nm has been selected.

Turn the instrument to single beam mode. The red indicator light on the S/D Beam button should be off. Single Beam mode is required for the light throughput optimization steps which follow.

Use the coarse wavelength setting to move the monochromator to the wavelength of the element line that was read ealier from the chart.

There is also a fine adjust that we will use to exactly position the wavelength. Although the wavelength scale is fairly accurate, we cannot just set it to the nominal wavelength. We will have to "tweak" the fine wavelength control a bit until the energy response of the instrument is a maximum.


The energy meter probably shows little or no response, even if the wavelength has been set correctly. We must first tell the instrument to adjust the electronic gain so we can see the signal. This is done with the EHT button.

Press the EHT button, then press Zero. This should move the energy meter to about one-third of full scale.

Turn the Fine Wavelength control until the energy meter shows a maximum reading. If the meter goes off scale, repeat the EHT - Zero operation.

With the meter at a maximum you should be sitting directly over the analysis wavelength.


Now adjust the lamp positioning controls until the energy meter shows a maximum. Use EHT-Zero to keep the meter on scale.

The light throughput should now be optimum. Do a final press of EHT - Zero for best electronic gain.


Put a piece of white paper directly over the burner slot and look for the position of the lamp beam. You must adjust the postion of the burner until the beam goes directly over the slot.

The burner positioning controls are located at the base of the burner assembly. Adjust In-Out and burner Rotation until the beam is lined up along the entire slot. Then ajust the burner height until the beam is about 1 cm above the slot.

Turn the instrument to Double Beam Mode (red light is On). You are now ready to light the flame and begin analyses.


Go on to Part 2: Lighting the Flame.
For questions or comments contact Dr. David L. Zellmer at david_zellmer@csufresno.edu
Last updated: 12/18/96