Operating the GBC 902
Atomic Absorption Spectrometer


Part 2: Lighting the Flame

David L. Zellmer
Department of Chemistry
California State University, Fresno
Fall 1996

WARNING!
Operation of this instrument involves the use of potentially explosive gases. DO NOT operate this instrument unless directly supervised by a faculty member or trained member of the staff. The directions below contain a great deal of safety information, but are not meant as a substitute for direct supervision.

Air is the oxidant for the flame. We use filtered house air in our lab. The pressure must be between 32 and 65 psi for proper operation.

You should not have to adjust this regulator as it is left on all of the time.


Turn the main valve of the acetylene tank on. Observe that the main tank pressure is well above 75 psi. If the tank pressure is near or below 75 psi, have the technician replace the tank with a fresh one. It is dangerous to use an acetylene tank below 75 psi because of the way the gas is stored in a solvent within the tank.

If the main tank pressure is well above 75 psi, turn the regulator valve clockwise until the outlet pressure shows 12 psi.

Tap on the gauge glass frequently as you turn up the pressure, since the needle tends to stick a bit.


Turn the oxidizer control to the AIR position. Observe that the air flow meter reads 6 or greater. You should not have to adjust the air flow control to get this flow rate since the manufacturer has preset this flow rate as a minimum value.

Check again that the hood has been turned on. We will add the acetylene and light the flame in the next few steps.


Turn the fuel control valve until the flow meter reads between 2 and 3.

Press and hold the Ignite button until the flame lights.

If anything does not seem right to you, turn the flame off by turning the fuel flow control clockwise until the needle valve seats. With the fuel choked off, the flame will go out.


Adjust the Fuel control until you have the type of flame called for in your analysis. If Fuel-Rich or Fuel-Lean is not specified, then turn down the fuel until the yellow glowing carbon particles just vanish from the flame.

This will be the Stoichiometric setting.

Reducing the fuel flow past the Stoichiometric point will produce a Fuel-Lean flame.


If you have not already done so, place the inlet tube for the nebulizer into a container of dust-free distilled water.

Aspirating room air is a bad idea, since the dust in the air can clog the very small inlet to the nebulizer.

You are now ready to begin analyses of your samples and standards.


Remember, if you need to turn the flame off, do so by turnining off its FUEL. Rotate the fuel control clockwise until the valve seats. The flame will quietly go out.

After the flame is out you can turn the air control off as well.



Return to Part 1: Setting up the Instrument.
Go on to Part 3: Analyzing the Sample.
For questions or comments contact Dr. David L. Zellmer at david_zellmer@csufresno.edu
Last updated: 12/18/96