Operating
the GBC 902 |
David L. Zellmer |
Operation of this instrument involves the use of potentially explosive gases. DO NOT operate this instrument unless directly supervised by a faculty member or trained member of the staff. The directions below contain a great deal of safety information, but are not meant as a substitute for direct supervision. |
When you are through running all of your samples make sure to aspirate plenty
of distilled water through the nebulizer to remove all traces of your samples
from the instrument before shutting it down.
|
|
To
turn the flame off, just turn off the FUEL by turning the FUEL valve clockwise
until it seats. The flame will go out quietly.
You can then turn off the Air using the oxidizer master control as shown.
|
|
Turn
off the acetylene at the main tank valve.
Even though the acetylene is now off, some gas still remains in the regulator
and the lines. We will now purge that gas from the system.
|
|
Turn
the oxidizer control back to AIR. Then turn up the FUEL control to allow the
acetylene to bleed from the lines.
The hood must remain on during these procedures.
|
|
Watch
the tank gauge and the output regulator gauge until they both read zero.
The lines are now purged.
|
|
Turn
the regulator valve counter-clockwise (screwing it OUT) until you feel it
become loose.
It is now in the zero pressure position.
|
|
Turn
the master oxidizer control from AIR to OFF.
Turn the FUEL control clockwise until it gently seats.
The gases are now all off.
|
|
Finally,
turn off power to the instrument.
If you have correctly followed all of these shutdown procedures, you will not have left any nasty surprises for the next person to use the instrument.
Please turn off the hood (unless someone else is using it) and clean up the
area before you leave.
|
|